LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 60-03-75
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 04/28/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump console.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative confirmed the reported malfunction but was unable to identify the root cause.Several parts were replaced and the issue persisted.The unit was returned to livanova (b)(4) for further investigation.Visual inspection did not identify any abnormalities or defects related to the reported issue.During functional testing and hardware analysis, the reported issue was reproduced and the root cause was traced to a defective component on the backplane board.The backplane board was replaced and a battery test was successfully completed before returning the unit to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event Description
Livanova (b)(4) received a report that the touchscreen of the centrifugal pump console was not illuminated and displayed no information during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP CONSOLE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6586595
MDR Text Key75817746
Report Number9611109-2017-00375
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K020571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 05/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-03-75
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/28/2017
Initial Date FDA Received05/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-