Model Number SLIMLINE 200¿ REUSABLE FIBER
Device Problem Material Fragmentation (1261)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/19/2015
Event Type  Injury  
Manufacturer Narrative
As part of a remedial activity, lumenis conducted a retrospective review of all its slimline 200u fiber complaint files from january 2015 to may 2017.An investigation of the reported event found that the reported malfunction of the slimline 200u fiber was similar to a device malfunction that was alleged to have caused or contributed to a potentially harmful situation.Because the company is aware that this malfunction was alleged to have caused or continued to a serious injury (reference mdr #3004135191-2017-00002), this event represents a reportable malfunction.Although the doctor was able to remove the broken fiber fragments from within the kidney and no patient harm had been reported, lumenis is also reporting this as an adverse event as medical intervention was required to preclude serious injury.
 
Event Description
A user facility reported that during a laser lithotripsy procedure the tip of a lumenis slimline 200u fiber broke off inside the patient and the broken pieces were retrieved successfully.The case was completed using a second slimline 200u fiber with no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE EZ 200
Type of Device
LASER FIBER DELIVERY DEVICE
Manufacturer (Section D)
LUMENIS LTD.
6 hakidma street
po box 240
yokneam, 20692
IS  20692
Manufacturer (Section G)
LUMENIS LTD.
6 hakidma street
po box 240
yokneam, 20692
IS   20692
Manufacturer Contact
alan vaisman
6 hakidma street
po box 240
yokneam, 20692
IS   20692
49599044
MDR Report Key6586897
MDR Text Key75815746
Report Number3004135191-2017-00053
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140388
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/10/2019
Device Model NumberSLIMLINE 200¿ REUSABLE FIBER
Device Catalogue Number0641-020-01
Device Lot Number0033931114
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/19/2015
Initial Date FDA Received05/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-