Catalog Number BI70002000
Device Problem Bent (1059)
Patient Problem No Patient Involvement (2645)
Event Date 04/27/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Onsite investigation was preformed and the field representative confirmed that the x-stage cover was damaged.Replacement of the cover resolved the issue.No further issues were reported.Replaced cover was not returned to manufacturer for analysis, therefore, no findings are possible.A full imaging system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A medtronic representative reported that the imaging system's x-stage cover was physically bent after the system lost its homing position.There was no patient present when this issue was identified.
 
Manufacturer Narrative
The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key6588399
MDR Text Key75872333
Report Number1723170-2017-02151
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/27/2017
Initial Date FDA Received05/24/2017
Supplement Dates Manufacturer Received06/30/2017
Supplement Dates FDA Received07/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/06/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-