Model Number SC-2316-50E |
Device Problem
Material Separation (1562)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/01/2017 |
Event Type
Injury
|
Event Description
|
A report was received that the patient's contacts came off of the distal tip of the lead.Fluoroscopy confirmed that contacts are in the subcutaneous tissue and would have to be removed by the surgeon.
|
|
Manufacturer Narrative
|
Sc-2316-50e (b)(4): device evaluation indicated that the complaint was confirmed.Visual inspection revealed distal electrodes #1-#3 were fractured and torn off.Cables are exposed at the separated section of the lead.Electrodes 1-3 were not returned.The cause of the lead damage couldn't be determined.Review of the device history record revealed no anomalies.
|
|
Event Description
|
A report was received that the patient's contacts came off of the distal tip of the lead.Fluoroscopy confirmed that contacts are in the subcutaneous tissue and would have to be removed by the surgeon.
|
|
Search Alerts/Recalls
|