Catalog Number 328438
Device Problems Break (1069); Device Or Device Fragments Location Unknown (2590)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 05/02/2017
Event Type  Injury  
Manufacturer Narrative
Results: a sample is not available for evaluation.A review of the device history record revealed one quality notification (b)(4) for defective material found during manufacturing of the reported lot # 6278672.This incident was unrelated to the customer's reported defect.Conclusion: without a sample, an absolute root cause for this incident cannot be determined as bd was not able to duplicate or confirm the customer¿s indicated failure mode.(b)(4).
 
Event Description
It was reported that a bd insulin syringe with the bd ultra-fine¿ needle 0.3ml 31g x 5/16" broke off in a consumer's stomach during an injection.The consumer went to an emergency department and received x-rays but the broken needle was not found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE 0.3ML 31G X 5/16"
Type of Device
INSULIN SYRINGE AND NEEDLE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key6592663
MDR Text Key76012719
Report Number1920898-2017-00064
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Patient
Type of Report Initial
Report Date 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328438
Device Lot Number6278672
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/02/2017
Initial Date FDA Received05/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-