Model Number SC-2218-50 |
Device Problems
Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 05/05/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model#: sc-2218-50 serial#: (b)(4) description: linear st lead, 50cm model#: sc-4316 lot#: 15187817 description:next generation anchor kit-sterile.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation and high impedances were noted.The patient underwent a revision procedure wherein the leads and clik anchor were replaced.No device malfunction was suspected but lead fracture was suspected.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Sc-2218-50 (sn: (b)(4)) device evaluation indicated that the complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that all of the cables were completely broken at the bent/kinked location of the lead.The bent/kinked location is 1 cm from the set screw mark of the clik anchor.There are no exposed cables at the fracture locations.Sc-2218-50 (sn (b)(4)) device evaluation indicated that the complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that multiple of the cables were completely broken at the bent/kinked location of the lead.The bent/kinked locations are 1 cm from both sides of the set screw mark of the clik anchor.There are no exposed cables at the fracture location.Sc-4316 (ln: 15187817) device evaluation revealed both anchors have one eyelet torn through.The small piece of silicone was not returned.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation and high impedances were noted.The patient underwent a revision procedure wherein the leads and clik anchor were replaced.No device malfunction was suspected but lead fracture was suspected.The patient was doing well postoperatively.
|
|
Search Alerts/Recalls
|