Model Number 9207010/57063
Device Problems Crack (1135); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2017
Event Type  malfunction  
Manufacturer Narrative
The customer was provided with information for purchasing a new power supply and the original power supply was requested for evaluation; however, the customer disposed of the power supply.This issue with a damaged rev n power supply for the pump in style device was addressed in investigation (b)(4) which was trended as part of the effectiveness check for (b)(4), which found that the power supplies are being damaged during shipment from the manufacturer to medela.This damage is causing the plastic housing to fail prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the power supply to medela was not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process was modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength was also been increased to further protect the power supply during shipping.
 
Event Description
On (b)(6) 2017, the customer alleged to customer service that her pump in style advanced breast pump will not power on with the power supply.The housing of the power supply has a crack in it and she can see into it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED - ON-THE-GO TOTE
Type of Device
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate drive
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate drive
mchenry IL 60050
Manufacturer Contact
robert sokolowski
1101 corporate drive
mchenry, IL 60050
MDR Report Key6597437
MDR Text Key76436858
Report Number1419937-2017-00154
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9207010/57063
Device Catalogue Number9207010/57063
Device Lot NumberREV N / UNK
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/08/2017
Initial Date Manufacturer Received 05/08/2017
Initial Date FDA Received05/30/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-