Catalog Number BI70002000 |
Device Problems
Break (1069); Difficult to Open or Close (2921)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/01/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.A medtronic representative went to the site to test the equipment.The representative reported that the winch assembly on the imaging system gantry was damaged due to shear and thus replaced.Additionally, the door cable slides were damaged and replaced by the medtronic representative.The imaging system then passed the system checkout and was found to be fully functional.The suspect winch and slides were returned to the manufacturer for evaluation.Both components failed visual inspection due to damaged components.
|
|
Event Description
|
A site radiologic technologist (rt) reported that, while in a spinal fusion, the imaging system gantry door would not open when prompted by the user.The site attempted to manually open the door, but a loud noise was heard emitting from the imaging system.The reported issue occurred when the site was done using the imaging system in the procedure.The site then removed the patient from the imaging system gantry area following the procedure.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no impact on patient outcome.No additional information was provided.
|
|
Manufacturer Narrative
|
The suspect cover was returned to the manufacturer for evaluation.Visual inspection found scratches on the cover, however the reported issue could not be replicated.
|
|
Search Alerts/Recalls
|