Model Number 37800
Device Problem Bent (1059)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/04/2017
Event Type  malfunction  
Manufacturer Narrative
Other applicable components are: product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2017 explanted: product type lead product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2017 explanted: product type lead.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A manufacturing representative (rep) reported on behalf of a healthcare provider (hcp) that during implant both leads were placed and the hcp pulled the first lead out of the abdomen through the laparoscopic port.The metal tip was pulled through the port at a slight angle and once it came through it was noticed that the metal tip was bent.The hcp was able to insert it into the header block of the battery without issue.Impedance was normal at approximately 500.They completed the case without any additional issues.Each lead impedance was checked and both c<(>&<)>2 and c<(>&<)>3 were below 500.The rep noted that they were not sure which lead was bent.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6600541
MDR Text Key76276009
Report Number3004209178-2017-11324
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169360686
UDI-Public00643169360686
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2017
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/04/2017
Initial Date FDA Received05/31/2017
Supplement Dates Manufacturer Received05/04/2017
Supplement Dates FDA Received10/04/2017
Date Device Manufactured02/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
-
-