MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/02/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Onsite investigation was preformed and the umbilical cable was found damaged, the upper latch portion of the connector was broken off.Replacement of the cable resolved the issue.No further issues were reported.Analysis of the returned cable confirmed the physical damage since one of the pins that hold the cable to the system was broken off, as a result, the cable could not stay connected to the system.A full imaging system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A medtronic representative reported that the site physically damaged the umbilical cable causing the mobile view station (mvs) to lose communication with the image acquisition system (ias).There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
sam saleh
attn:product quality experienc
826 coal creek circle
louisville, CO 80027
MDR Report Key6601401
MDR Text Key76299021
Report Number1723170-2017-02277
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2017
Initial Date FDA Received05/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-