NIHON KOHDEN CORPORATION ZM-521PA; TRANSMITTER Back to Search Results
Model Number ZM-531PA
Device Problems Overheating of Device (1437); Failure to Power Up (1476)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 05/03/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the transmitter over heats and will not turn on.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the transmitter over heats and will not turn on.
 
Manufacturer Narrative
The customer reported that the transmitter over heats and will not turn on.The unit was sent in to nihon kohden for evaluation.The batteries required for this investigation were not returned with the unit.During the evaluation, new batteries were inserted into the unit and heating could not be duplicated.Inspection of the negative contacts shows resin melting at the spring which per an nkc investigation on a similar incident is indicative of improper battery insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-521PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6602566
MDR Text Key76498511
Report Number8030229-2017-00158
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921115084
UDI-Public4931921115084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/31/2017,05/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-531PA
Device Catalogue NumberZM-531PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Report Sent to FDA05/31/2017
Distributor Facility Aware Date05/03/2017
Device Age49 MO
Event Location Hospital
Date Report to Manufacturer05/31/2017
Initial Date Manufacturer Received 05/31/2017
Initial Date FDA Received05/31/2017
Supplement Dates Manufacturer Received05/31/2017
Supplement Dates FDA Received10/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-