Model Number SIPAP
Device Problems Moisture Damage (1405); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Results of investigation: the carefusion factory service center received the suspect device, a sipap ventilator, and evaluated the device.An evaluation of the component confirmed the reported issue and found the power supply pcb defective.The factory service center repaired the ventilator and returned it to specifications.
 
Event Description
The customer reported that the pcb's fuse was blown and there were humidity/fluid issues on the back of the board.There was no patient involvement associated with this reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION
Type of Device
VENTILATOR, NON-CONTINUOUS
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6605213
MDR Text Key76593288
Report Number2021710-2017-06041
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
K031745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIPAP
Device Catalogue Number675-CFG-004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/03/2017
Initial Date FDA Received06/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-