• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. GAMMA SYSTEM - ENDO FEMORAL HEAD 12 / 14 TAPER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. GAMMA SYSTEM - ENDO FEMORAL HEAD 12 / 14 TAPER; PROSTHESIS, HIP Back to Search Results
Catalog Number 00781804300
Device Problem Component or Accessory Incompatibility (2897)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a total hip arthroplasty the femoral head could not be reduced into the compatible sized acetabular cup.The procedure was completed with a smaller size femoral head.No adverse event was reported as a result of this malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.As returned, the outside diameter was found to be conforming to print specifications.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAMMA SYSTEM - ENDO FEMORAL HEAD 12 / 14 TAPER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6607507
MDR Text Key76509955
Report Number0002648920-2017-00372
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PK955473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number00781804300
Device Lot Number37221291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/03/2017
Initial Date FDA Received06/02/2017
Supplement Dates Manufacturer ReceivedNot provided
04/04/2018
Supplement Dates FDA Received06/29/2017
04/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-