Catalog Number 00781804300 |
Device Problem
Component or Accessory Incompatibility (2897)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/01/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a total hip arthroplasty the femoral head could not be reduced into the compatible sized acetabular cup.The procedure was completed with a smaller size femoral head.No adverse event was reported as a result of this malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.As returned, the outside diameter was found to be conforming to print specifications.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|