• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM Back to Search Results
Model Number 81000
Device Problems Component Falling (1105); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported that during a procedure, the iv pole on the trima equipment unexpectedly lowers down.The customer stated that no adverse event occurred and no medical intervention was required.Information of patient (donor) or operator of the device is not known at this time.
 
Manufacturer Narrative
An internal report indicates no further related issues have been reported.One year of service history was reviewed for this device with one problem identified related to the reported condition with an incident date of (b)(6) 2017 (reported on 1722028-2017-00215).Correction: trima field action 30 has been initiated to notify all trima users to use precaution while transporting the device and a caution statement was included in the operator's manual.Corrective action: an internal capa has been initiated to evaluate reports of the iv pole dropping down suddenly.Root cause: the root cause of this failure was the iv pole fell down during a procedure.
 
Event Description
The customer declined to provide patient (donor) information and operator information since no adverse event occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
steve kern
10810 w. collins ave
lakewood, CO 80215
3032392246
MDR Report Key6607721
MDR Text Key76727396
Report Number1722028-2017-00200
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
BK040069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date 06/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number81000
Other Device ID Number05020583810006
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2017
Initial Date FDA Received06/02/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/07/2008
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-