Model Number 81000 |
Device Problems
Component Falling (1105); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/02/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation is in process.A follow-up report will be provided.
|
|
Event Description
|
The customer reported that during a procedure, the iv pole on the trima equipment unexpectedly lowers down.The customer stated that no adverse event occurred and no medical intervention was required.Information of patient (donor) or operator of the device is not known at this time.
|
|
Manufacturer Narrative
|
An internal report indicates no further related issues have been reported.One year of service history was reviewed for this device with one problem identified related to the reported condition with an incident date of (b)(6) 2017 (reported on 1722028-2017-00215).Correction: trima field action 30 has been initiated to notify all trima users to use precaution while transporting the device and a caution statement was included in the operator's manual.Corrective action: an internal capa has been initiated to evaluate reports of the iv pole dropping down suddenly.Root cause: the root cause of this failure was the iv pole fell down during a procedure.
|
|
Event Description
|
The customer declined to provide patient (donor) information and operator information since no adverse event occurred.
|
|