Model Number 16413 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/10/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Evaluation is in progress, but not yet concluded.The fsr troubleshot the problem and determined the temperature/pressure printed circuit board (pcba) was at fault.The fsr replaced the temperature/pressure pcba.The unit operated to manufacturer's specifications.The temperature/pressure pcba was returned to the manufacturer for further evaluation.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the channel a of arterial monitor would not display temperature.It would only display dashes.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.During laboratory analysis, product surveillance technician (pst) observed that a damaged input connector caused the channel a input to not display temperature.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|