• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problems Failure to Power Up (1476); Failure to Pump (1502)
Patient Problem No Patient Involvement (2645)
Event Date 05/08/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The investigation is on-going.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that a s5 roller pump did not turn on during setup.There was no patient involvement.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was unable to reproduce the reported issue.The pump was tested and was found to be functioning correctly.A serial readout was performed and sent to livanova (b)(4) for analysis.Analysis of the readout identified recorded alarms due to ¿cover open¿ and ¿master/slave¿ settings.It was suggested to the service representative that the pump cover be replaced, and the cardioplegia master/slave settings be removed.The service representative returned to the facility to replace the pump cover and found that it was broken at the hinge area.The service representative found that the cardioplegia pump was set as slave to the arterial pump, so if the arterial pump stopped the cardioplegia pump would also stop and not pull air across the membrane.The setting was not removed, as the setup was how the perfusionist wanted it.The service representative cleared the nvmem on the pumps and tested according to the service manual.The system was found to be working properly and the unit was returned to service.No further occurrences have been reported.As the issue could not be reproduced, a root cause was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich,
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6609906
MDR Text Key76627962
Report Number9611109-2017-00399
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2017
Initial Date FDA Received06/02/2017
Supplement Dates Manufacturer Received08/18/2017
Supplement Dates FDA Received09/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-