• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 8LPC
Device Problem Air Leak (1008)
Patient Problem Extubate (2402)
Event Type  Injury  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that the tracheostomy cuff would not hold air.Re-incubation of a replacement tube was required.
 
Manufacturer Narrative
One sample was received for analysis and the reported issue was confirmed.A inflation/deflation test was performed and it was observed the cuff deflated immediately.A visual inspection was performed and it was observed that the cuff was cut.Manufacturing controls are in place to detect cuffs with damage and to reduce the potential for occurrence during the manufacturing process.Information has been added to the database and trends will continue to be monitored.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Type of Device
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
sharon murphy
15 hampshire street
mansfield, MA 02048
2034925267
MDR Report Key6611856
MDR Text Key76668222
Report Number2936999-2017-05238
Device Sequence Number1
Product Code JOH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8LPC
Device Catalogue Number8LPC
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/09/2017
Initial Date FDA Received06/05/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/23/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-