• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. BASE PLATE 25 MM POST LENGTH +2 MM; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. BASE PLATE 25 MM POST LENGTH +2 MM; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Ossification (1428); Unspecified Infection (1930); Necrosis (1971); Pain (1994); Joint Dislocation (2374); Limited Mobility Of The Implanted Joint (2671)
Event Date 08/26/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).The customer has indicated that the product will not be returned to zimmer biomet for investigation, as the devices remain implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Concomitant devices: trabecular metal reverse polyethylene liner, catalog #: 00434903603, lot #: 62249294; trabecular metal reverse glenosphere 36mm diameter, catalog #: 00434903611, lot #: 62239886; trabecular metal reverse humeral stem 14mm diameter, catalog #: 00434901413, lot #: 62151241; inverse/reverse screw system, catalog #: 01.04223.042, lot #: 2564705; inverse/reverse screw system, catalog #: 01.04223.042, lot #: 2665729.This report is number 4 of 4 mdrs filed for the same patient (reference 0001822565-2017-03905 / 03907 / 03908).
 
Event Description
It is reported that the patient has been experiencing pain, recurrent dislocations and avascular necrosis following shoulder arthroplasty.No additional patient consequences have been reported.Additional information is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action(s) is/are required.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It is reported that the patient has been experiencing pain, recurrent dislocations, limited range of motion, and avascular necrosis following reverse shoulder arthroplasty.Patient alleges that the poly liner is wearing out, but x-rays showed no concerns.No additional patient consequences have been reported.Additional information is unavailable at this time.
 
Manufacturer Narrative
Upon receipt of information and reassessment of the reported event, it was determined that this device did not contribute to the event and should not have been reported.The initial report was forwarded in error and should be voided.
 
Manufacturer Narrative
(b)(4).The following report is submitted to relay additional information x-rays dated (b)(6) 2013 are reviewed and reports identified that there has been joint arthroplasty with essentially anatomic alignment and post operative changes.It was also noted in op-notes that the shoulder was reduced and carried through a range of motion; it was stable completing the right reverse shoulder replacement.No issues were noted.Previous investigation remains unchanged.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
(b)(4).Poly liner plus 3 mm offset 36 mm diameter cat# 00434903603, lot# 62249294.Glenosphere 36 mm diameter cat# 00434903611, lot# 62239886.Humeral stem 14 mm stem diameter 130 mm stem length cat# 00434901413, lot# 62151241.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2017-03905, 0001822565-2017-03907, 0001822565-2017-03908.
 
Event Description
It was reported that a patient underwent a revision procedure approximately 8 years postimplantation due to pain, osteolysis, poly wear and debris, poor bone quality, limited rom, dislocation, necrosis and possible infection.Due to the poor bone quality of the patient and humeral fracture, only a hemi revision was possible with the glenosphere being revised.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Review of the sterile cert indicates the device was sterilized per validated requirements.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: initial op notes and pathology report on (b)(6) 2013 noted preoperative diagnosis: right shoulder rotator cuff tear arthroplasty, failed hemiarthroplasty, biceps tendinitis and tearing with instability of biceps tendon.Previous incision utilized, encountered subdeltoid adhesions.Subscapularis was torn and incised off the anterior shoulder.Chronic tearing of supraspinatus and infraspinatus.Loose humerus hardware and suture removed.Bicep¿s sheath was transferred to pectoralis sheath.Cultures obtained from humeral canal.Labrum and biceps were excised.Drain placed.No complications.Portion of bone with fibrocartilage showing changes of degenerative osteoarthritis.The underlying marrow shows small areas of active hematopoiesis and focal osteopenia.Mri report on (b)(6) 2018 noted alignment is normal, small effusion with debris, minimal heterotopic ossification, particulate synovitis, 0.9 x 1 x 1.2 cm cyst area in the inferior glenoid (may be degenerative, erosion, or focal osteolysis), rotator cuff with no tendon retraction, mild to moderate atrophy and fatty replacement, tendon appears to attach from the lesser tuberosity and scarred to the pseudocapsule, stress reaction at the glenoid and humeral stem.No revision op notes were provided.The root cause of the reported event for possible infection was determined to be unrelated to the implanted zb devices.The root cause for the other reported issues remains unknown.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE PLATE 25 MM POST LENGTH +2 MM
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6618617
MDR Text Key76868455
Report Number0001822565-2017-03909
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model NumberN/A
Device Catalogue Number00434902502
Device Lot Number62301785
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/11/2017
Initial Date FDA Received06/07/2017
Supplement Dates Manufacturer Received10/09/2017
12/04/2017
07/03/2018
08/26/2021
08/05/2022
Supplement Dates FDA Received10/13/2017
12/04/2017
08/03/2018
03/19/2022
08/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.; SEE H10 NARRATIVE.; SEE H10 NARRATIVE.; SEE H10 NARRATIVE.; SEE H10 NARRATIVE.; SEE H10.
Patient Outcome(s) Hospitalization; Other; Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight110 KG
-
-