Boston scientific received information that one day post-implant this pacemaker and right atrial (ra) lead exhibited high out-of-range pace impedance measurements and loss of capture (loc) with only some sensing observed in unipolar configuration.The lead was believed to be underinserted and it was recommended that a procedure be performed to reinsert the lead.The physician elected to not revise the lead and instead program the patient's device to right ventricular therapy only.No adverse patient effects were reported.This system remains in service.
|