MAUDE Adverse Event Report: TERUMO CORPORATION IMUFLEX BLOOD BAG SYSTEM; IMUFLEX WB-RP BLOOD BAG SYSTEM WITH LR FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION IMUFLEX BLOOD BAG SYSTEM; IMUFLEX WB-RP BLOOD BAG SYSTEM WITH LR FILTER Back to Search Results
Catalog Number 1BBWGQ506A2
Device Problems High Test Results (2457); Filtration Problem (2941)
Patient Problem No Patient Involvement (2645)
Event Date 05/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation is in process.A follow up report will be provided.
 
Event Description
The customer reported elevated white blood cell (wbc) content in a filtered whole blood unit.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #:(b)(6).
 
Manufacturer Narrative
This report is being filed to provide additional information.Investigation: one set of blood bags was received for evaluation.The leukoreduction filter was tested for flow rate and air leaks.A slow flow rate was noted and it was confirmed there were no air leaks.The manufacturing records, test records, and inspection records were reviewed for abnormalities and none were found.The records regarding the particulate removal rates of the filter membranes were reviewed.All membranes conformed to established specification.Root cause:a definitive root cause for the observed elevated wbc count remains undetermined at this time.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the whole blood product could be donor-related (characteristics of blood).It also cannot be ruled out that a filtering error or other process error (the physical stress on filter membranes is greater than what is expected, trapped leukocytes are pushed out of the filter membranes) could have contributed to the higher-than-expected wbc content in the whole blood product.The instructions for use provide a caution to not squeeze or apply pressure to the filter while it is attached to the bag containing the filtered blood and to clamp the blood filled tubing before blood enters the filter in order to avoid leukocyte leakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMUFLEX BLOOD BAG SYSTEM
Type of Device
IMUFLEX WB-RP BLOOD BAG SYSTEM WITH LR FILTER
Manufacturer (Section D)
TERUMO CORPORATION
fujinomiya
Manufacturer Contact
steve kern
10810 w. collins ave.
lakewood, CO 80215
3032392246
MDR Report Key6621987
MDR Text Key77499913
Report Number9681839-2017-00088
Device Sequence Number1
Product Code CAK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/09/2017,06/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Catalogue Number1BBWGQ506A2
Device Lot Number161121KK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2017
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date05/17/2017
Device Age8 MO
Event Location Hospital
Date Report to Manufacturer06/09/2017
Initial Date Manufacturer Received 05/17/2017
Initial Date FDA Received06/08/2017
Supplement Dates Manufacturer Received06/28/2017
Supplement Dates FDA Received06/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-