Model Number 1095060
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/01/2017
Event Type  malfunction  
Event Description
It was reported that the power cord of this device was ripping and falling apart.On return of the device for evaluation, it was apparent that the insulation on the mains cord was degraded to the extent that bare wires were visible.Our evaluation concluded that in all likelihood the cord has been subjected to considerable external force.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ESSENCE
Type of Device
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
Manufacturer Contact
alessandro agosti
chichester business park
city fields way, tangmere
chichester, west sussex PO20 -2FT
UK   PO20 2FT
704231549
MDR Report Key6629122
MDR Text Key77235216
Report Number9681154-2017-00008
Device Sequence Number1
Product Code CAF
UDI-Device Identifier00383730001814
UDI-Public0383730001814
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042655
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1095060
Device Catalogue Number1099966
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/24/2017
Initial Date FDA Received06/09/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-