Catalog Number 9734403
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/12/2017
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2017 a replacement probe was shipped.Replacement probe resolved the issue.No further issues were reported.The suspected probe was returned to manufacturer for evaluation.Testing found that the tip of the probe was twisted severely.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
A medtronic representative reported that during a spinal fusion procedure with navigation system a navlock thoracic probe was found to have bent tip.There was no delay to the procedure.The procedure was completed with the use of navigation system.No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACIC PROBE TIP
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key6630225
MDR Text Key77429814
Report Number1723170-2017-02415
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00613994724014
UDI-Public00613994724014
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734403
Device Lot Number111102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/12/2017
Initial Date FDA Received06/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-