• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MANUF. SOLUTIONS SA PALINDROME; CATHETER, HEMODIALYSIS, IMPLANTED, COATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MANUF. SOLUTIONS SA PALINDROME; CATHETER, HEMODIALYSIS, IMPLANTED, COATED Back to Search Results
Model Number 8888145044P
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 2017/june/12.An investigation is currently underway; upon completion the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2017 that a customer had an issue with a dialysis catheter.The customer states there is a leak in the catheter under the luer connector.
 
Manufacturer Narrative
This complaint has not been confirmed.The actual sample involved in the reported incident was not returned for evaluations.No additional information, pictures or videos were received.Since no sample was returned for examination, it was not possible to evaluate it as part of a comprehensive failure investigation.A device history record (dhr) review revealed no discrepancies that may have contributed to a complaint of this failure mode.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements.In addition, all dhrs are reviewed for accuracy prior to product release.No additional information was received for investigation.If the sample is returned in the future, this complaint will be re-opened for further investigation.The available information was analyzed and it did not allow confirming a root cause for the event, however, process and design failure mode and effects analysis ( pfmea and dfmea) were reviewed in order to identify the possible causes for the failure.The potential causes were identifi ed as: defective material, operator failed to follow process and inspection procedures, customer misuse, or equipment malfunction.No trends or triggers have been found, therefore, a corrective and preventive action (capa) is not deemed necessary at this time.It must be noted that in-process controls, such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling, are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer states there is a leak in the catheter under the luer connector.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME
Type of Device
CATHETER, HEMODIALYSIS, IMPLANTED, COATED
Manufacturer (Section D)
COVIDIEN MANUF. SOLUTIONS SA
edificio 820 calle#2
zona franca coyol
alajuela 20101
Manufacturer (Section G)
COVIDIEN MANUF. SOLUTIONS SA
edificio 820 calle#2
zona franca coyol
alajuela 20101
Manufacturer Contact
edward almeida
15 hampshire st.
mansfield, MA 02048
5084524151
MDR Report Key6633276
MDR Text Key77306602
Report Number3009211636-2017-05164
Device Sequence Number1
Product Code NYU
UDI-Device Identifier10884521158092
UDI-Public10884521158092
Combination Product (y/n)N
Reporter Country CodeCI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/24/2018
Device Model Number8888145044P
Device Catalogue Number8888145044P
Device Lot Number1520400091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/25/2017
Initial Date FDA Received06/12/2017
Supplement Dates Manufacturer ReceivedNot provided
06/21/2017
Supplement Dates FDA Received06/21/2017
09/07/2017
Date Device Manufactured07/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-