Model Number ROTAFLOW ENGLISH/US
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Battery failure during transport.The customer reported that this issue was corrected prior to service technician maquet(stm) arrival.The customer found battery connector to be loose and reconnected battery connector.Customer reported no adverse effects to the patient.He performed functional and safety test and returned to clinical service.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.A supplemental medwatch will be submitted after new information has been received.
 
Event Description
(b)(4).During patient transport inside of the hospital the pump of the rotaflow would not operate in battery mode.The device was not changed out to another one.No adverse effects to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6633524
MDR Text Key77520041
Report Number8010762-2017-00190
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW ENGLISH/US
Device Catalogue Number701051712
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2017
Initial Date FDA Received06/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-