Model Number 72202976-
Device Problems Bent (1059); Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/15/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during device testing, the scope was bent and dislocating from seams.No patient injury was noted.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection was performed and showed a dented outertube, the outertube is jolted, laser welding damage, and the stopcock is damaged and leaky.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all manufacturers¿ quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.The instruction for use recommends ¿careful inspection of the operative hysteroscope before and after the procedure for possible signs of damage.Immediate detection and repair of minor damage will extend the life of the operative hysteroscope.¿ damage may occur if the hysteroscope is handled roughly during use.Excessive forces may result in traumatic insertion, chipping particles of the optical glass, or other damage to the unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR HANDPIECE
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG - OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
Manufacturer (Section G)
MITG - OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
Manufacturer Contact
sharon murphy
60 middletown avenue
north haven, CT 06473
2034925267
MDR Report Key6637358
MDR Text Key77450874
Report Number1643264-2017-21044
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202976-
Device Catalogue Number72202976-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/16/2017
Initial Date FDA Received06/13/2017
Supplement Dates Manufacturer Received12/14/2017
Supplement Dates FDA Received01/11/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-