Lot review: a review of the mfg.Lot database for lot# 3256398 (mfg.Date 06/2016) shows (b)(4) units were mfgd., tested, inspected, and released; cites no exception documents.Visual inspection: unit was visually examined no defect was found.Each lumen was observed to have no occlusions or anomalies.Functional testing: the temperature thermistor functioned along with the catheter being able to be calibrated per the dfu.Final analysis summary: the reported complaint could not be determined as it was unclear why the unit was being returned.No defect was found with the catheter.The catheter was able to be calibrated and the thermistor functioned.No failures were observed with any of the lumens.Icu medical wil monitor and trend.
|