Model Number 7209808-
Device Problems Difficult to Insert (1316); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2017
Event Type  malfunction  
Event Description
According to the reporter, during device testing, the unit was jammed and something seems to stuck inside connection port.There was no patient involvement noted.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Complaint of broken piece from mdu stuck in control unit's receptacle was confirmed.It appears that extreme force was used to remove mdu connector on power cord from hand piece receptacle on control unit which caused shaft to break off mdu connector and become stuck inside of hand piece receptacle.The complaint investigation has concluded that this is a user error since product has been in use at customer site for almost 3 years before incident occurred.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures; therefore, a device history record will not be performed.The root cause of the observed condition was determined to be a result of a misuse of the product.The ifu recommends ¿¿ careful inspection of the operative hysteroscope before and after the procedure for possible signs of damage.Immediate detection and repair of minor damage will extend the life of the operative hysteroscope.¿ damage may occur if the user attempts to connect a damaged or defective hand piece to the control unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG-OKLAHOMA CITY
75 s. meridianave
oklahoma OK 73107
Manufacturer (Section G)
MITG-OKLAHOMA CITY
75 s. meridianave
oklahoma OK 73107
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key6641151
MDR Text Key77610422
Report Number1643264-2017-21046
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209808-
Device Catalogue Number7209808-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/19/2017
Initial Date FDA Received06/14/2017
Supplement Dates Manufacturer Received11/03/2017
Supplement Dates FDA Received12/01/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-