Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Complaint of broken piece from mdu stuck in control unit's receptacle was confirmed.It appears that extreme force was used to remove mdu connector on power cord from hand piece receptacle on control unit which caused shaft to break off mdu connector and become stuck inside of hand piece receptacle.The complaint investigation has concluded that this is a user error since product has been in use at customer site for almost 3 years before incident occurred.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures; therefore, a device history record will not be performed.The root cause of the observed condition was determined to be a result of a misuse of the product.The ifu recommends ¿¿ careful inspection of the operative hysteroscope before and after the procedure for possible signs of damage.Immediate detection and repair of minor damage will extend the life of the operative hysteroscope.¿ damage may occur if the user attempts to connect a damaged or defective hand piece to the control unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|