Catalog Number 80440
Device Problems Incorrect Or Inadequate Test Results (2456); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Patient Involvement (2645)
Event Date 03/13/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The signals in the rdf indicate that the higher-than-expected wbc content in the platelet product was likely a result of an escape of wbcs from the lrs chamber during a portion of the procedure.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Weight was included due to limitations within internal system.Donor unit #: (b)(6).The platelet collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: the signals in the rdf indicate that the higher-than-expected wbc content in the platelet product was likely a result of an escape of wbcs from the lrs chamber during a portion of the procedure.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL ENHANCED PLATELET, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
steve kern
10810 w. collins ave.
lakewood, CO 80215
3032392246
MDR Report Key6647041
MDR Text Key77927521
Report Number1722028-2017-00242
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK150321
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2018
Device Catalogue Number80440
Device Lot Number12Z4112
Other Device ID Number05020583804401
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/22/2017
Initial Date FDA Received06/16/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight81
-
-