Discordant d-dimer results were obtained on 2 samples from the same patient on the sysmex cs-2100i system using the innovance d-dimer kit.The operator ran a patient sample for d-dimer (ddi) on the sysmex cs-2100i system and obtained a flagged ddi result 7.12 mg/l.The result was flagged with hemolytic sample and lipemic sample.This result was reported to the physician.The operator re-ran the same patient sample on the same sysmex cs-2100i system and obtained a ddi result of 7.39 mg/l.The result was flagged with error messages: hemolytic sample, lipemic sample, and defective sample volume.The instrument automatically diluted the same patient sample by 1/8 and a non-numerical (---.--) flagged result was obtained.The 1/8 diluted patient sample was flagged with error messages: hemolytic sample, lipemic sample, defective sample volume, dilution ratio was changed, and extrapolation boundary over (lower).The "extrapolation boundary over (lower)" flag indicates a result below clinical cut-off (i.E.<0.19 mg/l).The operator manually diluted the same patient sample by ½ and re-ran the sample on the same sysmex cs-2100i system.A ddi result of 1.96 mg/l was obtained.The patient blood was re-drawn to eliminate any pre-analytical issue with the original specimen and re-run on the same sysmex cs-2100i system.A flagged ddi result of 10.45 mg/l was obtained for this re-drawn patient sample.The result was flagged with error messages: hemolytic sample, lipemic sample, and defective sample volume.The instrument automatically diluted the same re-drawn patient sample by 1/8 and a non-numerical (---.--) flagged ddi result was obtained.The 1/8 diluted patient sample was flagged with error messages: hemolytic sample, lipemic sample, defective sample volume, dilution ratio was changed, and extrapolation boundary over (lower).The operator stated that the specimens were correctly filled and the volume flags are only seen because the operator re-ran the specimen multiple times.Both, original and re-drawn, samples have hemolytic flags but the operator stated that both samples did not appear hemolytic nor lipemic.Only the initial ddi result of 7.39 mg/l was reported to the physician.The correct result is unknown.There are no known reports of patient intervention or adverse health consequences due to the discordant d-dimer results.
|