• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+ USA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+ USA Back to Search Results
Catalog Number 10379675
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/25/2017
Event Type  malfunction  
Manufacturer Narrative
The dipping technique and cleaning procedure were reviewed with the customer and it was determined that the proper dipping technique and cleaning procedures were not being followed.The event has occurred due to operator error.The instrument is performing as intended.
 
Event Description
The customer reported discrepant results for urine leukocytes and blood between the clinitek status+ and the visual read of the multistix 10-sg strips.There was no injury reported due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+ USA
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key6650765
MDR Text Key77935964
Report Number3002637618-2017-00100
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/25/2017
Initial Date FDA Received06/19/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-