• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL CLUSTER HOLE SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL CLUSTER HOLE SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 502-11-56F
Device Problem Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2017
Event Type  malfunction  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Additional information has been requested and if received, will be provided in the supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
 
Event Description
The company representative reported that the surgeon had an issue with a trident hemispherical shell and the trident impactor.The cup was impacted into the patient and the surgeon couldn't unscrew the cup from the impactor.This meant the cup had to be removed and a new cup opened and new impactor used.The surgeon couldn't remove the cup even when it was removed from the patient.The stryker sales rep was handed the impactor and shell after the case and did manage to unscrew them, but it was very tight as if it was cold welded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT HEMISPHERICAL CLUSTER HOLE SHELL
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
brian lauro
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6652035
MDR Text Key77918149
Report Number0002249697-2017-01927
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K143085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial
Report Date 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2020
Device Catalogue Number502-11-56F
Device Lot Number51584101
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/22/2017
Initial Date FDA Received06/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-