Model Number URF-V2
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.The evaluation found the scope had an abnormal bending section during angulations.The bending section cover was removed and found the skeleton of the bending section broken on the insertion tube side.Based on the oem investigation finding; the investigations have determined that excessive torque loaded to the bending section through usage.The reported phenomenon could be reproduced when angulated the bending tube and the insertion tube is forcibly rotated 180 degrees at the same time.The instruction manual warns user "do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.".
 
Event Description
Olympus was informed that the tip of the scope was noted to be bent during reprocessing.It was reported that the device had been used in a procedure prior without issue.There was no patient injury reported.
 
Manufacturer Narrative
This supplemental report is being submitted to correct the report type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
408935-512
MDR Report Key6652275
MDR Text Key77938373
Report Number2951238-2017-00419
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Device Catalogue NumberURF-V2
Other Device ID Number04953170343582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2017
Initial Date FDA Received06/19/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/20/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-