The customer stated that they received erroneous results for one patient sample tested for the elecsys anti-tshr immunoassay (anti-tshr) on an unknown roche analyzer.It was asked, but it is not known if any erroneous results were reported outside of the laboratory.The sample initially resulted as > 40 iu/l.The sample was diluted and repeated, resulting as < 0.3 iu/l.The sample was then repeated without dilution, resulting as 4.0 iu/l.No adverse events were alleged to have occurred with the patient.The analyzer model and serial number were asked for, but not provided.The customer declined to provide any further information.A specific root cause could not be determined based on the provided information.Additional information required for the investigation was requested, but not provided.A general reagent issue can most likely be excluded.Possible root causes relate to sample quality and the presence of foam/bubbles on the reagent surface.
|