• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ANTI-TSHR, ANTIBODIES TO TSH RECEPTOR; ANTI-TSHR IMMUNOASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ANTI-TSHR, ANTIBODIES TO TSH RECEPTOR; ANTI-TSHR IMMUNOASSAY Back to Search Results
Catalog Number 04388780190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/06/2017
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).(b)(4).
 
Event Description
The customer stated that they received erroneous results for one patient sample tested for the elecsys anti-tshr immunoassay (anti-tshr) on an unknown roche analyzer.It was asked, but it is not known if any erroneous results were reported outside of the laboratory.The sample initially resulted as > 40 iu/l.The sample was diluted and repeated, resulting as < 0.3 iu/l.The sample was then repeated without dilution, resulting as 4.0 iu/l.No adverse events were alleged to have occurred with the patient.The analyzer model and serial number were asked for, but not provided.The customer declined to provide any further information.A specific root cause could not be determined based on the provided information.Additional information required for the investigation was requested, but not provided.A general reagent issue can most likely be excluded.Possible root causes relate to sample quality and the presence of foam/bubbles on the reagent surface.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTI-TSHR, ANTIBODIES TO TSH RECEPTOR
Type of Device
ANTI-TSHR IMMUNOASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6657281
MDR Text Key78258868
Report Number1823260-2017-01296
Device Sequence Number1
Product Code JZO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04388780190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/07/2017
Initial Date FDA Received06/21/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-