STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO13.2
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Code Available (3191)
Event Date 08/31/2016
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 13.2mm vticmo13.2 implantable collamer lens, -16.0/1.5/74 diopter, in the patient's right eye (od) on (b)(6) 2016.In the same surgery, the lens was explanted due to lens tear/beak during injection into the eye.On (b)(6) 2016 the lens was exchanged for the same size and diopter lens.The problem was resolved.The customer stated after loading the lens, it was not put in bss.The customer though that is the reason causing the lens tore, and the surgeon has been trained again.As of the date of mdr submission, the lens has not been returned.
 
Manufacturer Narrative
Device evaluation: the product was returned dry, in a micro centrifuge vial with residue on product.Visual inspection found the lens cut in two pieces, missing a piece of haptic and residue on the lens surface.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
MDR Report Key6657575
MDR Text Key78089712
Report Number2023826-2017-01019
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date05/31/2019
Device Model NumberVTICMO13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/26/2017
Initial Date FDA Received06/21/2017
Supplement Dates Manufacturer Received03/20/2018
Supplement Dates FDA Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age32 YR
-
-