The manufacturer representative (rep) reported that the healthcare provider (hcp) was in surgery on the day of the report to replace the implantable neurostimulator (ins).The hcp told the rep that the impedance was over 800 ohms, but it was not greater than 4,000 ohms.It was also indicated that the hcp wanted to know if the ins could be activated, and that they may go back in at a later date to revise the lead.The rep mentioned that the hcp made sure the connections were wiped off and secure.There were no symptoms or complications reported as a result of this event.
|