The device was not returned; however, photographs were received and evaluated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection of the photographs showed evidence of broken tubing.It is unclear regarding the exact location on the device where the tubing is broken off because the multirate control module is not shown in the photos.The reported condition was verified, however the cause could not be determined from the images.Should additional relevant information become available, a supplemental report will be submitted.
|