The cause for the elevated replicates of advia centaur xp ee2 is unknown.The customer noted that there were no related errors found with the system when the ee2 tests were run and the system maintenance is performed.Siemens continues to investigate.The performing quality control/taking corrective action section of the instructions for use states: "if the quality control results do not fall within the expected values or within the laboratory's established values, do not report results.Take the following actions: -verify that the materials are not expired.-verify that required maintenance was performed.-verify that the assay was performed according to the instructions for use.-rerun the assay with fresh quality control samples.-if necessary, contact your local technical support provider or distributor.".
|
Mdr 1219913-2017-00145 was filed on june 27, 2017 reporting elevated advia centaur xp enhanced estradiol (ee2) results for four replicates of a quality control using one lot of cuvettes compared to the quality control results obtained with a different lot of cuvettes.July 5, 2017 - additional information.Lot number, gtin, date of manufacture and expiration date of the advia centaur enhanced estradiol were not available at the time of the initial filing.Lot number: 31977037, gtin 00630414573342 date of manufacture: 12/12/2016, expiration date: 10/12/2017.The customer confirmed that the cuvettes were stored according to the manufacturer's storage temperature recommendations of -20c to 60c.Siemens is unable to investigate further because the cuvettes are no longer available.Several factors can attribute to imprecision with the estradiol assay including reagents, calibrations, control integrity, instrument performance.Siemens is unable to confirm that cuvettes are the cause of the imprecision seen at this site.
|