• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752022
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/09/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
An ophthalmic surgeon reported that the cutter would not pull out from the trocar during a procedure.The procedure was completed without replacing the product.No patient harm reported.A product sample has been requested, however, it has not been received for evaluation at the manufacturing site.
 
Manufacturer Narrative
No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.The trocar cannula and probe samples were visually inspected.The cannula was then dimensionally inspected for inner diameter and was found conforming.The probe needle body, middle, and tip were measured and deemed conforming.The probe needle was fit tested for function into a conforming trocar and the fit was deemed conforming.The probe and cannula testing met specification.The exact root cause of the probe could pull probe from trocar cannot be determined from this evaluation.No specific action with regard to this complaint was taken by the manufacturing location because the probe was manufactured to its specification.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6672503
MDR Text Key78576923
Report Number1644019-2017-00542
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752022
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/13/2017
Initial Date FDA Received06/27/2017
Supplement Dates Manufacturer Received09/11/2017
Supplement Dates FDA Received09/17/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-