Model Number 980
Device Problems Occlusion Within Device (1423); Noise, Audible (3273)
Patient Problem No Code Available (3191)
Event Date 05/24/2017
Event Type  Injury  
Event Description
It was reported that, while in use on a patient, a 980 ventilator experienced a sever occlusion and generated a "loud popping" noise.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
 
Manufacturer Narrative
The service engineer (se) replaced the exhalation valve flow sensor (evq) and performed calibrations, extended self-testing and all tests passed.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer Contact
ray maroofian
2101 faraday ave
carlsbad, CA 92008
7606035334
MDR Report Key6672843
MDR Text Key78564465
Report Number8020893-2017-06759
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/30/2017
Initial Date FDA Received06/28/2017
Supplement Dates Manufacturer Received10/03/2017
Supplement Dates FDA Received10/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-