Model Number M00510880
Device Problem Separation Failure (2547)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/07/2017
Event Type  Injury  
Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.Device component code (b)(4) relates to device problem code (b)(4) for the reported event of tip won't detach.According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx basket was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, a trapezoid¿ rx basket was used in attempt to crush a stone.However, the basket was unable to crush the stone and the tip failed to detach.The papilla was dilated with a cre balloon.When one of the stones fell out of the basket, the basket was able to be successfully removed from the bile duct.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID¿ RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6673623
MDR Text Key78595644
Report Number3005099803-2017-01961
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296393
UDI-Public(01)08714729296393(17)20170830(10)19658785
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/30/2017
Device Model NumberM00510880
Device Catalogue Number1088
Device Lot Number19658785
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/09/2017
Initial Date FDA Received06/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-