Initial reporter phone (b)(6).(b)(4).Device evaluated by manufacturer: the pump, hypotube, shaft, and tip were microscopically and tactile inspected.Inspection revealed that the male connector that attaches to the pump was cracked/damaged.Functional testing was done by placing the device in the console, fluid was leaking from the crack in the connector and the device would not get into priming mode.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause has been determined to be related to a supplier manufacturing issue.(b)(4).
|
It was reported the device broke and stopped aspiration during the procedure.During the use of an angiojet® zelantedvt¿ catheter in a pulmonary embolism procedure, the device was set up properly as usual.After running the catheter for forty seconds it stopped working.At this point, after checking over the console and catheter, it was noticed the outlet adaptor nut on the pump assembly was cracked.The device was removed from the patient and the procedure was completed with manual aspiration with no consequences to the patient.
|