THERAKOS XTS PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Disconnection (1171); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.No kit lot was provided; therefore, no batch record review could be conducted.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).Device not returned.
 
Event Description
Customer called to report the tubing disconnected from the centrifuge bowl when attempting to empty residual fluid from the centrifuge bowl.The tubing disconnected and leaked after the procedure was completed.All blood products from the treatment and return bags were reinfused to the patient.The customer was unable to return the residual volume.Customer stated there were no alarms during the procedure.Customer will not be returning product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XTS PHOTOPHERESIS SYSTEM
Type of Device
XTS PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS
hampton NJ
Manufacturer (Section G)
THERAKOS, INC.
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
53 frontage road
p.o. box 9001
hampton, NJ 08827
MDR Report Key6676467
MDR Text Key78983134
Report Number2523595-2017-00135
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Model NumberNOT APPLICABLE
Device Catalogue NumberNOT PROVIDED
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2017
Initial Date FDA Received06/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-