• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO REVO MRI SURESCAN; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO REVO MRI SURESCAN; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE Back to Search Results
Model Number RVDR01
Device Problems Incorrect Measurement (1383); Device Displays Incorrect Message (2591); Reset Problem (3019); Missing Test Results (3267)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/27/2017
Event Type  Injury  
Event Description
It was reported that upon interrogation of an implantable pulse generator (ipg) there was no battery voltage measurement available.A bitflip was confirmed, the device was reprogrammed and remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the battery measurement was not available.Battery voltage not available due to bit flip in lead impedance/battery voltage trend data.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVO MRI SURESCAN
Type of Device
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6678563
MDR Text Key78764910
Report Number3004209178-2017-14024
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00613994614742
UDI-Public00613994614742
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/28/2014
Device Model NumberRVDR01
Device Catalogue NumberRVDR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2017
Initial Date FDA Received06/29/2017
Supplement Dates Manufacturer Received06/29/2017
06/29/2017
Supplement Dates FDA Received07/11/2017
10/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age8 YR
-
-