Catalog Number 1805034300
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/01/2017
Event Type  malfunction  
Event Description
It was reported that the comfort gel mattresses have been allegedly sliding/moving with patient on and off of the stretcher.There was patient involvement, however no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
It was identified that the customer may be tucking sheets underneath the support surface which resulted in reduced contact area between the bottom of the support surface and the support platform, which can cause the surface to slide.This issue was resolved for the customer by the explaining to the account manager the importance of proper sheet placement.
 
Event Description
It was reported that the comfortgel mattresses have been allegedly sliding/moving with patient on and off of the stretcher.There was patient involvement, however no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
Supplemental submitted to include udi.
 
Event Description
It was reported that the comfortgel mattresses have been allegedly sliding/moving with patient on and off of the stretcher.There was patient involvement, however no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFGEL SE W/O FIRE BAR, 30IN
Type of Device
MATTRESS, FLOTATION THERAPY, NON-POWERED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
kristen canter
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key6681388
MDR Text Key79078723
Report Number0001831750-2017-00284
Device Sequence Number1
Product Code IKY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number1805034300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2017
Initial Date FDA Received06/30/2017
Supplement Dates Manufacturer Received06/01/2017
06/01/2017
Supplement Dates FDA Received12/07/2017
03/26/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-