Catalog Number 1805034300 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/01/2017 |
Event Type
malfunction
|
Event Description
|
It was reported that the comfort gel mattresses have been allegedly sliding/moving with patient on and off of the stretcher.There was patient involvement, however no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
It was identified that the customer may be tucking sheets underneath the support surface which resulted in reduced contact area between the bottom of the support surface and the support platform, which can cause the surface to slide.This issue was resolved for the customer by the explaining to the account manager the importance of proper sheet placement.
|
|
Event Description
|
It was reported that the comfortgel mattresses have been allegedly sliding/moving with patient on and off of the stretcher.There was patient involvement, however no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
Supplemental submitted to include udi.
|
|
Event Description
|
It was reported that the comfortgel mattresses have been allegedly sliding/moving with patient on and off of the stretcher.There was patient involvement, however no adverse consequence or clinically relevant delay in treatment was reported.
|
|