Investigation summary: the affect product was returned for evaluation and the event report was confirmed.The vital port was returned unassembled, along with the catheter lock and a length of catheter measuring approximately 45.5 cm.The visual inspection revealed there were no signs of holes or damage to the catheter when observed under magnification.The suture holes were not utilized.The instructions for use was reviewed which states under port suturing and site closure: "three sutures, with at least one at each end of the port, are recommended".A product evaluation was performed which showed the suture holes were not utilized during this case, and there was only one needle stick observed in the septum.Neither the port or the catheter were occluded.The customer supplied photos/images and there is evidence of redness of the implant area.The device history record was also reviewed for the lot number of the device, and no signs were found to indicate that a nonconforming product was released to the field.This device lot has not been previously reported on a complaint.A quality engineering risk assessment concluded that no risk reduction is necessary.
|