MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC.1818910 QUICKSET GRATER HANDLE ASSMB; HIP INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC.1818910 QUICKSET GRATER HANDLE ASSMB; HIP INSTRUMENT/TRIAL Back to Search Results
Catalog Number 244000520
Device Problem Difficult to Insert (1316)
Patient Problem No Information (3190)
Event Date 06/05/2017
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the grater handle that connects to power is stripped and it does not stay connected (power adapter was later used properly with a different grater handle).
 
Manufacturer Narrative
Examination of the returned instrument confirmed the reported observation.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET GRATER HANDLE ASSMB
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6685701
MDR Text Key79040067
Report Number1818910-2017-20562
Device Sequence Number1
Product Code HTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number244000520
Device Lot NumberA0708
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/05/2017
Initial Date FDA Received07/05/2017
Supplement Dates Manufacturer Received08/03/2017
Supplement Dates FDA Received08/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-