• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FREE PSA; TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FREE PSA; TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA Back to Search Results
Catalog Number 03289788190
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/21/2017
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).Unique identifier (udi)#: (b)(4).
 
Event Description
The customer stated that they received erroneous results for one patient sample tested for the elecsys free psa immunoassay (fpsa) and the elecsys total psa immunoassay (tpsa) on a cobas 8000 e 602 module (e602).The fpsa result was higher than the tpsa result.The same results occur when other blood samples of the patient are tested (the fpsa result is greater than the tpsa result).This medwatch will apply to the fpsa assay.Patient identifier (b)(6) for information related to the tpsa assay.The sample resulted as 4.78 ng/ml for fpsa and 0.327 ng/ml for tpsa.No adverse events were alleged to have occurred with the patient.The e602 analyzer serial number was asked for, but not provided.
 
Manufacturer Narrative
The patient sample was provided for investigation.The customer's results were confirmed during the investigation.The sample contained an interferent to the tpsa assay.This limitation is covered in product labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREE PSA
Type of Device
TEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6692359
MDR Text Key79526321
Report Number1823260-2017-01402
Device Sequence Number1
Product Code MTG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P000027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03289788190
Device Lot Number20987700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/21/2017
Initial Date FDA Received07/07/2017
Supplement Dates Manufacturer Received06/21/2017
Supplement Dates FDA Received07/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-