• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. MERLIN PROGRAMMER; PACEMAKER PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. MERLIN PROGRAMMER; PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2017
Event Type  malfunction  
Manufacturer Narrative
The result of the analysis confirmed the reported event of black screen at start up and a burning smell.Inverter and power cables of the display are burned and damages are extended until the pepin board.
 
Event Description
It was reported that the programmer screen would not turn on and the programmer smelled burned.There were no reported adverse patient consequences related to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6692401
MDR Text Key79245011
Report Number2017865-2017-06131
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Other Device ID Number05414734500661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2017
Initial Date FDA Received07/07/2017
Supplement Dates Manufacturer Received09/26/2017
Supplement Dates FDA Received09/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-