During a procedure regarding the gastroenterology, the ten nm-400u-0523s were used.It was reported that the needles could be extracted a little bit, but the needles could not be retracted into the sheath for all ten devices.The procedure was canceled.There was no patient injury reported.This report describes the first device.
|
This is a supplemental report for mfr report # 8010047-2017-00856 to provide additional information based on the evaluation result of the subject device.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.As a result of the evaluation on july 31, 2017, it was confirmed that the needle of the subject device could not be retracted into the sheath.The sheath of the subject device was buckled at 1055 mm from the distal end.The device history record for the lot indicated no abnormality with the event-related items.Based on the similar cases in the past, the needle might be unable to retract into the sheath due to the large friction between the sheath and the needle after the sheath was buckled.The sheath might be buckled, because excessive load was applied to the sheath when the subject device was inserted into the endoscope, when it was taken out from the sterile package, or when it was checked before use.The instruction manual of the device has already warned as follows: when inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
|