Model Number 840
Device Problem Occlusion Within Device (1423)
Patient Problem No Patient Involvement (2645)
Event Date 06/11/2017
Event Type  malfunction  
Event Description
It was reported that an 840 ventilator had a safety valve occluded.The ventilator was not in use on a patient at the time the of event.The service engineer (se) verified the issue and replaced the safety valve.The unit passed all testing and operated within the manufacturing specifications.
 
Manufacturer Narrative
A safety valve was returned to covidien/medtronic¿s failure analysis.An investigation was performed and the failure analysis technician reported that the reported condition could not be duplicated.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer Contact
ray maroofian
2101 faraday ave
carlsbad, CA 92008
7606035334
MDR Report Key6702022
MDR Text Key79649563
Report Number8020893-2017-06886
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number4-840120DIUU-US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/12/2017
Initial Date FDA Received07/11/2017
Supplement Dates Manufacturer Received07/28/2017
Supplement Dates FDA Received08/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-